Contact Us

MTEG Technische Import Und Export GmbH
Chamissoweg 5, D-63456 Hanau, Germany
Tel.: +49-6181-67130
Fax: +49-6181-671317
Email: info@mtegde.com

8400 Sodium Analyzer

8400 Sodium Analyzer

    8400 sodium analyzer is used for continuous on-line measurement of sodium ion in ultra pure water.It is widely
 used in power plants, semi-conductor industry, pharmaceutical industry and other industries.

MEASURE

Measurement range: 0 – 1000 μg/L; 0 – 60 ℃

Display resolution: 0.1 μg/L; 0.1 ℃

Precision: ± 5% of the measured value

Repeatability: ± 2% of the measured value

Temperature probe: NTC 2.252 k ohm

Automatic temperature compensation range: 0 – 60 ℃

Sensor inputs: Potential Signal

Input impedance: ≥ 3×1014Ω

OPERATING CONDITION

Ambient temperature: 0 – 50 ℃

Relative humidity: < 85%

Power supply: AC 220V ±22 V, 50 Hz ± 1Hz

Altitude: ≤ 2000 m

MEASURE CHAMBER / PROBE

Pressure: ≤ 0.5 MPa

Flow: 100 to 400 ml / min, stable

CALIBRATION

Calibration type: 2 points/ maual

ANALOGUE OUTPUT

Output signals: 1 galvanically outputs insulated

Type: 4 – 20 mA

Maximum load: 500 Ω

Accuracy: 0.05 mA

ALARMS

Alarm number: 2

Breaking power (ohmic overcharge): 240 V AC, 3 A max; 28 V DC, 6 A max

RS485

Baud rate: 4800, 9600, 115200

PROGRAMMING

Language: English

Display: 4.3″ Color LCD (480×272 pixels)

Memorizer: Data of 270 days (1 point / 5 mins)

MECHANICAL CHARACTERISTICS

Dimensions: Panel: 444 mm wide × 630 mm high × 183 mm deep

Weight: 15 kg (Panel)

Tightness: IP65