8400 Sodium Analyzer
8400 sodium analyzer is used for continuous on-line measurement of sodium ion in ultra pure water.It is widely
used in power plants, semi-conductor industry, pharmaceutical industry and other industries.
MEASURE
Measurement range: 0 – 1000 μg/L; 0 – 60 ℃
Display resolution: 0.1 μg/L; 0.1 ℃
Precision: ± 5% of the measured value
Repeatability: ± 2% of the measured value
Temperature probe: NTC 2.252 k ohm
Automatic temperature compensation range: 0 – 60 ℃
Sensor inputs: Potential Signal
Input impedance: ≥ 3×1014Ω
OPERATING CONDITION
Ambient temperature: 0 – 50 ℃
Relative humidity: < 85%
Power supply: AC 220V ±22 V, 50 Hz ± 1Hz
Altitude: ≤ 2000 m
MEASURE CHAMBER / PROBE
Pressure: ≤ 0.5 MPa
Flow: 100 to 400 ml / min, stable
CALIBRATION
Calibration type: 2 points/ maual
ANALOGUE OUTPUT
Output signals: 1 galvanically outputs insulated
Type: 4 – 20 mA
Maximum load: 500 Ω
Accuracy: 0.05 mA
ALARMS
Alarm number: 2
Breaking power (ohmic overcharge): 240 V AC, 3 A max; 28 V DC, 6 A max
RS485
Baud rate: 4800, 9600, 115200
PROGRAMMING
Language: English
Display: 4.3″ Color LCD (480×272 pixels)
Memorizer: Data of 270 days (1 point / 5 mins)
MECHANICAL CHARACTERISTICS
Dimensions: Panel: 444 mm wide × 630 mm high × 183 mm deep
Weight: 15 kg (Panel)
Tightness: IP65